TritonTM System Control Sofware
System automation
- Fully automatic cool-down/warm-up routines
- Fully automatic pre-cooling/mixture collection
- Sample change wizard for systems with load-locks
- Fully integrated magnet control, including vector rotate magnet software
Triton command interface
- Platform-independent remote scripting interface
- Standard internet (TCP/IP) protocol
- Compatible with any programming language that can access the TCP port e.g. LabVIEW / Matlab / Python / C / C++ / C# / Visual Basic / HP VEE etc.
- Monitoring and control of automatic software routines and system parameters
- Definable access levels (read only/read write etc.) with security
CLICK HERE to access your free software upgrade.
-
Experimental control
-
In-built safety features
-
System logging
-
User support
-
- Programmable temperature and field sequences
- Programming library supplied with example LabVIEW VIs
- Integration of dilution refrigerator/magnet control and experimental data acquisition
- Continuous monitoring of all pressure gauges
- System automatically enteres "safe state" if necessary
- Automatic software intervention in case of overpressure with fail-safe in-line pressure relief valves protecting the mixture and dilution refrigerator in the event of extended power failures
- UPS option - automatic alert emails can be sent to user in the event of a power failure
- Monitoring of magnet temperature during field sweep, high temperature control, sample loading
- Automatic magnet de-energization in event of power cut, pulse tube cooler failure, excessive heat load on cooler second stage
- Automatic logging of system parameters (temperature, sensor resistances, pressure readings, magnetic field and current, pump speed, 3He flow etc.)
- Data can be exported as an ASCII text file, compatible with Excel
- User-initiated and software-controlled events are logged
- Detailed remote interface
- Remote system access by Oxford Instruments support staff can be enabled to aid system diagnostics
- Free software upgrades available for download from Oxford Instruments' website
- All system settings are backed up prior to shipping from the factory