The Kelvinox®JT is a small, dipstick-style, dilution refrigerator which features a Joule-Thomson condensation stageKelvinoxJT dip-stick dilution refrigerator meaning the KelvinoxJT can operate in any 4 K environment – wet or dry – as it does not rely on a 1 K pot. KelvinoxJT is ideal for neutron or X-ray scattering experiments because there is no liquid helium in the sample horizontal plane.

The refrigerator is compatible with our Cryofree TeslatronPT, our wet Integra magnet systems, any liquid helium transport dewar or any VTI with a sample tube diameter of at least 50 mm.

  • Can operate in any liquid helium environment, any dry 4 K (exchange gas) environment or in any 50 mm VTI
  • Quick cool-down time: typically 6 hours from room temperature to base
  • 20 µW of cooling power at 100 mK
  • Base temperature in 4 K (exchange gas or liquid helium) environment: 40 mK
  • Base temperature in a VTI: 25 mK
  • Experimental space: inner diameter 43 mm x length 180 mm
  • Inner vacuum chamber (IVC) with automatic exchange gas control
  • The IVC is sealed using vacuum grease of caff paste (no Indium is required)
  • Wide temperature range: when operated in a VTI the temperature can be controlled between 25 mK and 300 K
  • One spare 6 mm line-of sight port for installing experimental wiring

System includes:

  • Automatic gas handling system with pneumatically operated valves
  • Automatic software routines including auto-condense, mixture removal and temperature control
  • Rack mounted, oil-free pumps
  • System computer with pre-installed control software
  • Pressure gauges for gas control
  • Gas tank with optimized 3He/4He mixture
  • Lakeshore thermometry read-out and temperature control electronics
  • Generically calibrated Ruthenium oxide (RuO2) sensors on the mixing chamber and on the still (fully calibrated sensors available on request)
  • Fully calibrated Cernox sensor at 4 K
  • Remote command interface through TCP/IP
  • Cold trap in liquid nitrogen dewar
  • All interconnecting lines
  • IVC sorb heater

Specifications:

Base temperature

≤ 50 mK

Base temperature stability

± 1 mK

Maximum temperature

30 K

Continuous operation

Standard

Sample environment

Vacuum
(No liquid Helium in horizontal sample path)

Cooling power at 100 mK

≥ 30 µW

Cooling power at 120 mK

≥ 30 µW

Sliding seal assembly

Included with system

3He / 4He mixture

Optional *

We are able to supply the 3He / 4He with the system. Customer can also use their own system.

Downloads & further Information

 

To download our Kelvinox product guide click here

To download our Kelvinox specification sheet click here

Related Products

Standard dilution refrigerator - KelvinoxMX

Standard dilution refrigerator - KelvinoxMX

Dilution refrigerator for multi experiments - KelvinoxMX

Product Support

Help Desk support

Help Desk support

A team is available to answer phone or email requests.

Installation and Service

Installation and Service

We can take care of the installation and maintenance of your product for complete piece of mind.

System Maintenance, Repair and Upgrades

System Maintenance, Repair and Upgrades

A full repair and maintenance facility is available; our dedicated technicians repair and service hundreds of systems per year.

Dedicated Spares and Accessories

Dedicated Spares and Accessories

Oxford Instruments Direct - Cryospares offers an extensive range of cryogenic spares, magnet spares, accessories and consumables - with a fast and efficient service, expert support and rapid delivery of products world-wide.

Extended Warranty

Extended Warranty

Extended warranty and PM including parts and labour. This contract allows you to safeguard your investment with complete peace of mind.

Mercury Support

Mercury Support

Support website for customers using MercuryiTC temperature controller and MercuryiPS magnet power supply.