Maximum magnetic field at sample position 5T
Magnetic field stability in persistent mode (switch included) 0.01 %/hr
Magnetic field homogeneity 2% over 5 mm diameter spherical volume
Maximum magnetic field sweep rate 1.0 T/min
Temperature range on the sample heat exchanger 6 K to 300 K measured with pressurised LHe
storage dewar (see note 1)
6 K achieved with RV12 pump (option)
Temperature range on a demountable copper
sample holder (standard)
7 K to 300 K measured with pressurised LHe
storage dewar (see note 1)
6 K achieved with RV12 pump (option)
Temperature stability ± 0.1 K measured over 10 minutes
Typical cool-down time from room temperature
to 6 K
4 hours
Typical average liquid helium consumption at 6 K 2 litres per hour
Typical liquid helium consumption per day 25 litres, based on 10 hours of operation and including
two complete sample changes

NOTE 1: Pressurisation of the liquid helium dewar is the responsibility of the customer and a check should be made to establish that this is permitted within their laboratory safety regulations.