Shasta 9700001 Shasta 9700003
Operating voltage range 0-50kV 0-60kV
Maximum power 50W 60W
Maximum beam current 1.0mA
DC Filament Supply Current: 0.3 to 3.5A
Voltage: 0 to 5.0 VDC
Voltage Regulation Load: < 0.01% for 50% of max load variation
Line: < 0.01% for 10% change in input voltage
Current Regulation Load: < ± 2µA (Beam Current)
Line: < ± 2µA (Beam Current)
Ripple < 100V peak to peak
Stability ± 0.1% over an 8-hour period after 30-minutes warm-up
Input Voltage & Power 24VDC, ± 10%; 100 Watts
Voltage Control Local: via multi-turn potentiometer (kV ADJ)
Remote: via external voltage source 0 to 10V (accuracy ± 1%)
Interlock Short to GND through a 12V lamp: HV/ON, OPEN:HV/OFF
Protection Over voltage, over current protection. Arc, short circuit
Temperature Conditions Operational: 0 to 45°C
Storage: -20 to +85°C
Temperature Coefficient 0.01% per °C, voltage and current
Dimensions 140mm H x 83.5mm W x 240mm D
5.5" H x 3.3" W x 9.45" D
Weight 3.6kg
7.9 lbs
Regulatory & Safety Meets the requirements of IEC61010-1:2010, EN61010-1:2010, UL61010-1:2012, CAN/CSA
C22.2 No. 61010- 1:2012 and 2006/95/EC Low Voltage Directive. Product carries the TUV SUD
c/us mark.